Subscribe to the SLI Medical mailing list to receive updates on new arrivals, special offers and our promotions.
The store will not work correctly in the case when cookies are disabled.
Get Up to 25% OFF PPE MasksGowns * Limited time only.
SLI MEDICAL IS A WHOLESALE SITE! WE ONLY SELL TO QUALIFIED FACILITIES (Such As Hospitals, Physician Offices, & Distributors). All Addresses are Verified.
Refurbished in an ISO 13485 Certified FDA registered Facility in an ISO 13485 Certified FDA registered Facility
Two filling options: Easy-fil and funnel fill
The easy-fil filler is designed to simplify agent filling and help minimize agent leaks while filling
Large diameter control dial which incorporates fine graduations of 0.2% between 0 and 1% and 0.5% from 1 to 8%
Ergonomic agent dial release on the Tec 7 vaporizer allows either left or right-hand operation
Calibration: The dials of the Tec 7 vaporizers for enflurane
isoflurane and halothane are calibrated to 5% v/v in steps of 0.2% up to 1% and in steps of 0.5% from 1% to 5% at 21°C using O2 at 5 L/min as the carrier gas
Calibration of Sevoflurane: The dial of the Tec 7 vaporizer for sevoflurane is calibrated to 8% v/v
dial is marked in steps of 0.2% up to 1% v/v and in steps of 1% between 1% and 8%
Resistance to gas flow: Off: Isolated
no resistance from vaporizer
On: 5 L/min O2: 10 to 15 cm H2O at 21°C ±2°C
Accuracy: 5 L/min O2 @ 21°C ± 2°C: Vaporizer 5%: ±0.25% of delivered agent or ±15% of dial setting (whichever is greater)
Vaporizer 8%: ±0.4% of delivered agent or ±15% of dial setting (whichever is greater)
Flowrate: Vaporizers are designed to provide consistent output throughout the clinical flow range from 200 mL/min to 15 L/min